Enmet SDS-97D Manual de usuario Pagina 6

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 13
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 5
SDS – 97D Sensor / Transmitter ENMET Corporation
4
2.2.2 Sensor/Transmitter with ISB
Installation with a recommended Intrinsic Safety Barrier properly connected and grounded renders the sensor/transmitter
nonincendive as defined by the National Electrical Code. An MTL 787S+ barrier is recommended connected as shown in
Figure C. Connections at the sensor/transmitter are shown in Table 2 as follows:
Table 2: Wiring at S/T
Position Function
J3-1 V+, the power supply
J3-2 signal, the positive side of the loop
J3-3 ground, the negative side of the loop
J3-4 shield, only if the sensor/transmitter enclosure is not grounded
Figure C: Wiring to ISB
When the sensor/transmitter is mounted and wired, assure that the cover gasket is in place, replace the cover, and secure it with
the two cover screws.
Allow the sensor/transmitter to stabilize for 3 to 24 hours.
1 V+
2 Signal
3 Ground
4 Shield, if necessary
SDS
97D
Sensor / Transmitter
3
4
24 V
DC
Regulated
Power
Supply
Readout
4-20mA
Earth ground
J3
Safe Area
Hazardous Area
1
2
ISB, MTL 787S+
+
Vista de pagina 5
1 2 3 4 5 6 7 8 9 10 11 12 13

Comentarios a estos manuales

Sin comentarios