Sensor/Transmitter, Toxic Gas & Oxygen ENMET Corporation
5
4.0
Standard Ranges, Calibration Gases, Voltage Readings at Calibration Points and
Approximate Times to Sensor Stabilization
Table 1: Electrochemical Sensors
Gas
Symbol
Range*
(ppm)
Calibration Gas
Correlation
Reading
(mV)
Stabilize
(min.)
ENMET Gas
Part Number
Ammonia NH
3
0 – 100 25ppm NH
3
200 4 03218-025
Arsine (2) AsH
3
0 – 0.5 0.5ppm PH
3
0.375ppmAsH
3
400 3 03202-005
Boron Trichloride BCl
3
0 – 30 10ppm HCl 10ppm BCl
3
233 3 03202-010
Carbon Monoxide CO 0 – 100 50ppm CO 300 2 03219-050
Chlorine (2) Cl
2
0 – 10 5ppm Cl
2
300 4 03231-005
Chlorine Dioxide ClO
2
0 – 1.0 0.3ppm ClO
2
** 220 4 **
Dichlorosilane SiH
2
Cl
2
0 – 30 10ppm HCl 3ppm SiH
2
Cl
2
140 3 03202-010
Ethylene Oxide C
2
H
4
O 0 – 30 5ppm C
2
H
4
O 167 3 03232-005
Fluorine (2) F
2
0 – 10 3ppm Cl
2
7.5ppm F
2
400 3 03231-003
Hydrogen H
2
0 – 1000 800ppm H
2
420 3 03227-800
Hydrogen Bromide HBr 0 – 30 5ppm HCl 25ppm HBr 433 3 03202-005
Hydrogen Chloride (2) HCl 0 – 30 10ppm HCl 233 3 03202-010
Hydrogen Cyanide HCN 0 – 30 10ppm HCN 233 2 03203-010
Hydrogen Fluoride (1) HF 0 – 10 5ppm Cl
2
3ppm HF 220 4 03202-005
Hydrogen Sulfide H
2
S 0 – 100 20ppm H
2
S 180 2 03214-020
Nitrogen Dioxide NO
2
0 – 30 5ppm NO
2
167 2 03233-005
Nitric Oxide NO 0 – 100 25ppm NO 200 1.5 03234-025
Oxygen O
2
0 – 30 % 20.9% O
2
379 2 03296-209
Ozone (1) O
3
0 – 1.0 0.3ppm O
3
** 220 3 **
Phosgene (1) COCl
2
0 – 1.0
0.66ppm COCl
2
**
364 4 **
Phosphine (2) PH
3
0 – 1.0 0.5ppm PH
3
300 1.5 03205-005
Silane SiH
4
0 – 30 5 ppm SiH
4
167 2 03306-005
Silicon Tetrachloride SiCl
4
0 – 30 5ppm HCl 25ppm SiCl
4
433 3 03202-005
Sulfur Dioxide SO
2
0 – 30 10ppm SO
2
233 1.5 03215-010
N
OTE
: All specifications stated in this manual may change without notice.
(1) Require use of Sample Draw System.
(2) Recommended use of Sample Draw System, for enhanced sensitivity and speed of response.
*0 = 4mA = 100mV, upper point = 20mA = 500 mV
**Gas not available in cylinders, a gas generator must be used
N
OTE
: If a non-standard range and/or a non-standard calibration gas are used, utilize the equations in Section 3.0 to determine
voltage readings.