EX2000 ENMET Corporation
11
4.3 Replacing the EX-2000 Sensor
The EX-2000 is equipped with a catalytic sensor with a microprocessor source.
Being a vital element of the gas detector, the sensor module must be treated with care. Mechanical impacts, splashing with
water etc… may impair the quality of the measurement or even, in extreme cases destroy the sensor.
A water-repellent and anti-dust filer fixed to the sintered metal protects the head of the sensor.
To replace the sensor and filter:
Remove sensor cover by unscrewing the sensor cover from the instrument.
Note the placement of the cover seal, filter and sensor seal.
Unplug sensor and remove.
Plug-in the new sensor.
Replace sensor seal, filet and cover seal
Replace sensor cover and screw into instrument.
Figure 2: Sensor Located under the Sensor Cover
5.0 Replacement Part Numbers
5.0 Replacement Part Numbers5.0 Replacement Part Numbers
5.0 Replacement Part Numbers
ENMET replacement part numbers:
Description Part Number
Programming Plug 02552-005
Calibration cup 02552-004
Sensor 02552-001
Sensor Filter 02552-003
Battery pack with plug 02552-012
Charger, 110Vac 67051-050
Charger, 220Vac 67051-051
Sensor Cover
Filter
Sensor
Sensor Seal
EX-2000 Instrument