EX-5175-EC ENMET Corporation
10
Normal Display Mode
5ppm
MENU
H
OLD the magnet over the
M
ENU switch for 2 – 4 seconds
to enter the Maintenance Menus
Exit
S
ELECT
Tap the
S
ELECT switch to return to the Normal
Display Mode.
See Section 5.2.3
MENU
Tap the magnet over the MENU switch to cycle through Maintenance Menus
OR
Zero
Cal OK
If
the Zero signal is within Preset Specs the EX-5175-EC will
display Cal OK, See Section 5.2.1
If the Zero signal is not within Preset Specs the EX-5175-EC will
display Bad ZERO
NOTE: Some software revisions require the SELECT switch be
tapped to accept the signal.
S
ELECT
MENU
Bad ZERO
PV:
0
S
ELECT
Tap the SELECT switch to initiate Zero adjustment
Apply
Calibration Gas until signal value becomes
stable (about 1 to 4 minutes) See Figure 4
If cal is good display will indicate OK or Same
If cal is not within preset “range” display will
indicate Bad Sens
MENU
PV:
0
20
Cal OK
Same mV
Bad Sens
Span
S
ELECT
S
ELECT
OR
OR
NOTE: You can change the Calibration Gas Level.
HOLD the magnet over the SELECT switch
The M
ENU
switch changes digit indicated by underscore cursor
The S
ELECT
switch locks underscored digit and moves to next digit
See Section 5.2.2
S
ELECT
mA Span
MENU
S
ELECT
100
To change mA Span set point:
Tap the M
ENU
switch until mA Span is displayed
Tap the S
ELECT
switch to display the set point
The M
ENU
switch changes digit indicated by underscore cursor
The S
ELECT
switch locks underscored digit and moves to next digit
To
return to Normal Gas Display:
Tap MENU switch until EXIT is displayed
Then tap SELECT switch
Select
Menu
To change Alarm set points:
Tap Menu switch until Alarm to be changed is displayed
Tap Select switch to display the set point
The M
ENU
switch: changes digit indicated by underscore
cursor
The S
ELECT
switch: locks in the underscored digit and
moves to next digit
If change is not within range display returns to first digit
If change is within range display moves to Set Time Delay
Use M
ENU
and S
ELECT
switches as above to change time
delay. Between 0 and 5 seconds is allowed
If change is within range display moves to next menu
Λ - Indicates increasing alarm
V - Indicates decreasing alarm
SetTD
sec
0
Alarm3
S
ELECT
Λ
50
MENU
SetTD
sec
0
Alarm1
MENU
S
ELECT
Λ
10
SetTD
sec
0
Alarm2
MENU
S
ELECT
Λ
20
Figure 5: EX-5175-EC Maintenance Menu Flow chart