Enmet ISA-44-RALE-OD Manual de usuario Pagina 16

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 38
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 15
ENMET Corporation ISA-44RALE-OD/ISA-44RAHE-OD/ISA-44E-OD
12
Figure 4: Interior Terminal Enclosure
Table 3: Wiring Information Terminal Enclosure
For ISA-44RALE-OD / ISA-44E-OD (when Purge and Horn are supplied)
Terminal
Enclosure
Connection
Signal
TB5-6 MOS Sensor Ground
TB5-7 MOS Sensor Heater
TB5-8 MOS Sensor Signal
TB5-2 CO High Alarm Relay, Common
TB5-1 CO High Alarm Relay, Normally Open
TB6-1 CO High Alarm Relay, Normally Closed
TB5-5 CO Low Alarm Relay, Common
TB5-4 CO Low Alarm Relay, Normally Open
TB5-3 CO Low Alarm Relay, Normally Closed
TB6-2 Oxygen Alarm Relay, Common
TB6-3 Oxygen Alarm Relay, Normally Open
TB4-1 AC Ground
TB4-2 110 VAC (black)
TB4-3 110 VAC (white)
NOTE: The terms "normally open" and "normally closed" correspond to the de-energized positions of the relay contacts.
The relays are energized during normal (non-alarm) channel operation, so the "normally open" position is closed
and the "normally closed" position is open during non-alarm channel operation. The relays are de-energized
during alarms or when power to the channel is interrupted.
Terminal
Block 6
Terminal
Block 5
Terminal
Block 4
Vista de pagina 15
1 2 ... 11 12 13 14 15 16 17 18 19 20 21 ... 37 38

Comentarios a estos manuales

Sin comentarios