TX2000 / OX2000 ENMET Corporation
11
4.3.1A TX2000 Calibration Menu
Enter the Calibration Menu. First make zero adjustment.
1. Press the BACKLIGHT switch to toggle
from programming to calibration menu:
Switch to display the calibration menu
2. Press ENTER switch:
Enter the calibration menu
3
Display : zero adjustment
4. Press ENTER switch:
To enter the zero adjustment procedure
5.
Display the current value "zero".
6. Press UP or DOWN switches:
To adjust value to “zero” if necessary
7. Press ENTER switch.
To set the zero adjustment
Then make sensitivity adjustment.
Attach the calibration cover over the sensor.
Connect the hose from the calibration kit and inject the calibration gas ( >10% range of the gas detected
otherwise only the zero will be confirmed), maintaining a flow rate of 30 l/h.
Then continue scrolling the calibration menu:
1.
Display the adjustment of sensitivity
2. Press ENTER switch:
To enter the sensitivity adjustment procedure
3.
Example:
The display indicates proper calibration gas. Apply the gas:
Wait for the signal to stabilize.
4. Press UP or DOWN switches:
To adjust calibration gas correct value
5. Press ENTER switch:
To enter the adjustment of sensitivity
6. Display:
Display "request for confirmation"
6. Press ENTER switch.
Acceptance of request for confirmation
7. Press ALARM switch to toggle between
yes and no:
Display confirmation : NO
Display "yes or no" of confirmation
(with each depression of the button)
Display confirmation: YES
8. Press ENTER switch.
Enter the confirmation chosen
NO: The measurements are not entered
YES: The measurements are entered
9.
Return to normal display of current measurement
Note: When the calibration procedure has been completed, do not forget to remove the calibration
cap from instrument. Failure to do so will cause the instrument to be inaccurate.
PrG
– 0 –
CAL
– S –
100
CnF
No Yes