4.3 Additional Features
4.3.1 Data Retrieval
This function enables the user to view the last 10 manual samples. Use the download software to retrieve all data required.
Accessing the Data
1. Hold down the SAMPLE button and press the ON-OFF button once. Release the SAMPLE button.
2. The Instrument will briefly display: HEAt
The instrument will then flash between displaying the following two screens: Temperature reading and °C
3. Once the instrument is in this mode press the Cal ▲ and Cal ▼ buttons at the same time and release. The display will briefly
show: dAtA
4. The display will then flash between the following two screens (as an example): run0 and 0.08
The first screen indicates which previous sample is being displayed, the most recent first,
from run0 (most recent), to run9 (the oldest).
Use the Cal▲ and ▼ buttons to move to the desired sample reading.
By pressing the SAMPLE button when viewing a reading,
the instrument will display the reading in mg/m
3
, as shown below:
ALt0 and 0.56
Press the ON-OFF button when done to turn the instrument off.
4.3.2 Clearing the Non FRAM Memory
As with the older
htV
model, the
htV-m
has sufficient memory capacity
to log the last 10 samples as well as the FRAM memory.
While viewing the data, hold down both the Cal (▲ and ▼) buttons at the same time,
until the display has run through all the following screens:
CLr3 CLr2 CLr1 CLr0 [ ]
The non-FRAM memory will now be clear.