RECON/B SERIES ENMET
11
5.4 Sensor Replacement
Sensor life in the RECON/B SERIES is typically: 2 years for Carbon Monoxide & Hydrogen Sulfide and 1year for Oxygen and
Combustible.
The sensor needs to be replaced when the R
ECON/B SERIES can no longer be calibrated or if the sensor becomes erratic.
Contact ENMET for sensor replacement information.
Table1: RECON/B SERIES Factory Set Parameters
Model Gas Range Low Alarm High Alarm TWA STEL Cal Gas
RECON/NH3-B Ammonia 0-100ppm 25ppm 50ppm 25ppm 35ppm
RECON/CO-B Carbon Monoxide 0-1000ppm 35ppm 100ppm 25ppm 200ppm 500ppm
RECON/H2-B Hydrogen 0-1000ppm 35ppm 250ppm N/A N/A
RECON/HS-B Hydrogen Sulfide 0-100ppm 10ppm 15ppm 10ppm 15ppm 40ppm
RECON/O2-B Oxygen 0-30%vol 19.5%vol 23.5%vol N/A N/A 20.9%
RECON/SO-B Sulfur Dioxide 0-100ppm 2ppm 5ppm 2ppm 5ppm
RECON/EX-B Methane 0-100%LEL 10%LEL 50%LEL N/A N/A 50%LEL
RECON/EX-B Propane 0-100%LEL 10%LEL 50%LEL N/A N/A
6.0 Specifications and Troubleshooting Tips
6.1 Specifications
ELECTRICAL
POWER
Working Voltage
3.6V
DC Lithium battery,1200 mAh Rechargeable
Working time continuous, non-alarm conditions
Combustible gas: Approximately 8 hours
Toxic gas: Approximately 300 hours
Charge time 4 to 6 hours
OPERATION
Temperature:
–20C to +50C
Relative Humidity
5 to 95% RH, non-condensing
MECHANICAL
Dimensions:
2.4 x 3.9 x 1.3 inches(60 x 100 x 33 mm)
Weight:
5 oz. (140g) including battery
Ingress Protection IP 65
Approval Designed to be intrinsically safe
NOTE: All specifications stated in this manual may change without notice.
6.2 Troubleshooting Tips
MALFUNCTION POSSIBLE REASONS SOLUTIONS
Unable to Turn on the detector
low voltage Charge Battery
Detector Fault Contact ENMET
Circuit malfunction Contact ENMET
No reaction to the tested gas
Exhausted Sensor Replace Sensor
Circuit malfunction Contact ENMET
Inaccurate indication
Exhausted sensor Replace the sensor
Overdue Calibration Calibrate
Wrong time display
Exhausted battery Charge and reset the time
Electromagnetic disturb Reset the time
Zero calibration function failed Too much drift Calibrate or replace the sensor