3.1.1 Wiring the AM-5175
The electrical installation should conform to appropriate electrical codes, such as the National Electrical Code in the United
States.
W
ARNING
:
The compliance of the installation to appropriate codes is not ENMET’s responsibility.
The AM-5175 should be powered through circuit breakers provided for this purpose.
3.1.2 Power Supply
The input power can vary from 100 to 240 V
AC
, 50/60 Hz. Power should be connected to the Power Input Terminal TB1 and
the Ground screw. See Figure 4 for location. Instrument is supplied with a strain relief and a standard U.S. 3 prong power line
cord.
For DC wiring 24V
DC
may be wired to J12, (J12-1)position 1 + with ground connected to (J12-2)position 2.
Upon supplying power to the AM-5175:
The green power on LED is lit.
The display backlight is lit, and instrument will step through a start-up sequence: unit serial number and software revision
may be shown on the display.
The instrument may go into alarm briefly, but the sensors stabilize quickly. If the instrument persists in alarm, acknowledge the
alarm by pressing the S
ELECT
button. If alarm persists longer than 30 minutes, call ENMET customer service personnel.
W
ARNING
:
Continuous gas detection and alarm systems (110V
AC
/220V
AC
/ 24V
DC
/12V
DC
powered) become inoperative upon
loss of primary power. Contact factory for specifications and pricing of backup battery systems.
AC Power Supply Terminal: TB1
Label on PCB Function
110V
AC
TB1 ACN Neutral
TB1 ACL Line
Ground
Screw
AC GND
220V
AC
TB1 ACN Neutral
Optional TB1 ACL Line
Ground
Screw
AC GND
C
AUTION
: 110/220 V
AC
when applied
DC Power Supply/4-20mA
Terminal: J12
Position Function
1 +
24 V
DC
power
2 GND
3 4 - 20 mA out
Sensor Terminal: J8
Position Function Color
1 + Sensor Red
2 Sensor Signal White
3 Sensor Ground Black
Figure 4: Power Terminal Connections AM-5175
Cover Inside View
Opened Upward
Caution
: 110/220 Vac
when applied