5.2.2 Maintenance Menu Flow Chart
The maintenance menu diagram is shown in Figure 6 Maintenance Menu Flow Chart. From the operational display, press
the
OPTION
switch 6 times; "Enter MAINTENANCE Menu" is displayed.
Figure 6: GSM-60 Maintenance Menu Flow Chart.
O = Press Option
S = Press Select
MAINTENANCE MENU
Exit maint menu
PTION
to return to top of maintenance menu.
S
ELECT
to return to operational menu.
5.3.9 for instructions on how to exit Maintenance Menus.
Operational
From Operational Menu
Press
O
PTION
(6 times)
S
Enter Maint Menu
O
S
Enter Password
0000
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
O
(6)
O
S
MAINTENANCE MENU
Set New Password
Password
XXXX
O
S
Changes digit indicated by underscore cursor
Locks underscored digit and moves cursor to next digit
for changing password instructions.
O
S
Calibration
Select (Gas)
MAINTENANCE MENU
Calibration
PTION
5.3 for calibration instructions.
O
S
MAINTENANCE MENU
Set Alarm1
Select: XX
PTION
until the gas alarm to be
5.3.3 for setting alarms instructions.
O
S
Select: XX
MAINTENANCE MENU
Set Alarm2
PTION
until the gas alarm to be
5.3.3 for setting alarms instructions.
O
S
MAINTENANCE MENU
Set Alarm Delays
Select: XX
PTION
until the gas alarm delay to be
5.3.4 for setting alarms instructions.
S
MAINTENANCE MENU
Configure Alarms
R1 L
PTION
changes letter indicated by underscore cursor
5.3.5 for configuring relay instructions.
O
S
MAINTENANCE MENU
Relay Failsafes
R:1 Failsafe ON
PTION
changes Failsafe setting from ON to OFF
5.3.6 for configuring relay failsafe instructions.
MAINTENANCE MENU
Scale mA Xmtrs
O
PTION
until gas to be Set-Up is displayed
5.3.2 for transmitter set-up instructions.
Scale mA Xmtrs
Select (Gas)
S
O
S
MAINTENANCE MENU
mA Output Span
Select: (Gas)
PTION
5.3.7 for output span instructions.