Enmet MRI-5175 Manual de usuario

Busca en linea o descarga Manual de usuario para Instrumentos de medición Enmet MRI-5175. ENMET MRI-5175 User Manual Manual de usuario

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 2
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 0
ENMET Corporation
80003-503
Original April 2011
Page 1 of 2
Page 1 of 2
Guide for MRI-5175 Sensor Replacement
An Oxygen sensor must be replaced when it can no longer be calibrated. Typical sensor life is 3 to 5 years in a 20.9%
oxygen environment.
To replace a sensor, perform the following steps:
Turn off the electrical power.
Open the display panel and remove the four cover (lid) retention screws and remove the lid. See Figure 1.
Remove the old O
2
sensor, by unscrewing it, and replace it with a new sensor. See Figure 1.
Replace cover(lid) with retention screws.
Turn on the electrical power.
Wait 3 – 4 hours, then recalibrate the sensor per the instructions.
MRI-5175 Exterior View
(Optional)
V+
SIG
GND
MRI-5175 Lid, Interior View
Sensor Wiring
J8 Sensor Wires
V+
Not Used
Sig
Red wire
Gnd
Black wire
Figure 1: MRI-5175 Sensor Replacement
Figure 2: Calibration MRI-5175
O2 Sensor
Sensor Wires
To J8
Terminal J8
Power/Fault LED
MENU Switch
SELECT Sw
itch
Calibration Adapter
on Sensor
Menu
Select
Power/
Fault XX X
XXX
XXX
% O2 b y vol.
Cylinder,
Calibration Gas
Vista de pagina 0
1 2

Indice de contenidos

Pagina 1 - Sensor Wiring

ENMET Corporation 80003-503 Original April 2011 Page 1 of 2 Page 1 of 2 Guide for MRI-5175 Sensor Replacement An Oxygen sensor must be replaced when

Pagina 2 - Altitude Calibration

ENMET Corporation 80003-503 Original April 2011 Page 2 of 2 Page 2 of 2 Standard Calibration The following procedure is suitable for altitudes below

Comentarios a estos manuales

Sin comentarios