Step 4
Remove the sensor board from the rear of the sensor – do not disconnect the wires from the main board.
Step 5
Remove sensor insert using maintenance tool or slip joint pliers.
Step 6
Mark the sensor pin position & undo the snap ring that retains the sensor in its housing.
Step 7
Replace the sensor taking note of the pin marking in step 6.