EX-6100
ENMET Corporation
6
3.5 Following Completion of the Installation:
a) Replace the top display board using the locating pins on the underside of the board to ensure that it makes connection
with the remainder of the electronics - see Figure 8.
b) Ensure that the front cover has not been contaminated with dirt - paying particular attention to the thread. Replace the
cover by rotating it several times clockwise until it reaches its limit then secure using the set screw – do not over
tighten.
Figure 8: Replacing the Display Board
3.6 Applying power
Caution: The cover must be in place before applying power
Once the installation is complete and the covers are secure then power can be applied. The power source should be between 8
and 24 volts dc.
When power is first applied to the EX-6100, an initialization procedure is performed as follows:
All segments on the LCD are shown 8.8:8.8
The software version is displayed 1.3.0
The manufacturer code is displayed SSCL
The sensor type is displayed P
Warm up count down 30
The detected gas level present is displayed and the EX-6100 becomes operational.
3.7 Indicators
Three LED indicators are visible from the front panel of the instrument. These are:
Red Alarm Illuminates when either alarm level has been reached.
Green Power Always ON when power is connected.
Yellow Key press Illuminates when the magnetic switches are operated.