Enmet SPECTRUM ON-LINE Manual de usuario Pagina 24

  • Descarga
  • Añadir a mis manuales
  • Imprimir
  • Pagina
    / 30
  • Tabla de contenidos
  • MARCADORES
  • Valorado. / 5. Basado en revisión del cliente
Vista de pagina 23
ENMET Corporation Spectrum ON-LINE
20
SPECTRUM Series for Sulphur Dioxide(4S sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
carbon monoxide 300 <3
hydrogen sulphide 15 0
nitric oxide 35 0
nitrogen dioxide 5 -5
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor Concentration in ppm
chlorine 1
hydrogen 200
hydrogen cyanide 10
hydrogen chloride 5
ethylene 100
SPECTRUM Series for Arsine (2E sensor)
The following gases cause a reading:
Interferant
Gas/Vapor
Interferant Concentration
in ppm
Instrument Reading
in ppm
phosphine 100 ppb 110
chlorine 5 -400
hydrogen cyanide 10 100
ammonia 100 10
diborane 100 35
silane 1 100
germane 1.1 100
hydrogene selenide 50 ppb 5
The following gases or vapors have been tested and do not cause a reading:
Interferant Gas/Vapor Concentration in ppm
hydrogen sulfide 1*
sulfur dioxide 2
hydrochloric acid 5
carbon monoxide 300
carbon dioxide 5000
nitrogen 100 %
hydrocarbons % range
hydrogen 1000
*with onboard filter; continuous high level exposure may reduce the filter efficiency
Vista de pagina 23
1 2 ... 19 20 21 22 23 24 25 26 27 28 29 30

Comentarios a estos manuales

Sin comentarios