ENMET Corporation OMNI-4000 S/N 6705 and above
54
3.8 Replacing a Toxic Gas Sensor
This procedure is used for the installation of a new toxic gas sensor.
Equipment Required
A small phillips head screwdriver
A new toxic gas sensor
Replacing the Toxic Gas Sensor
Use the screwdriver to remove the cover which contains the sensor openings
Remove the old toxic gas sensor
Install the new toxic gas sensor. The connector ensures a correct fitting. Be careful not to damage the male part
of the connector.
Put the cover protecting the sensors back into place
Possible locations of toxic sensors are channels 2, 3, and 4. See Figure 3-14.
Figure 3-14: Acceptable Locations of Oxygen and Toxic Sensors
Checking the Toxic Gas Sensor
In the interests of safety, check the toxic gas sensor by applying a calibration gas of known contents. To do this:
Turn the OMNI-4000 on by pressing the ON/OFF switch
Using the calibration cover, apply the gas as during a calibration
Check that the concentration of the calibration gas from the cylinder corresponds to the display for the appropriate
channel
Remove the gas and the calibration cover
Explosive Gas
Sensor
CO2 Sensor
(if needed)
Toxic Gas or
Oxygen Sensor
Locations